    The wireless probe design of the ValProbe® system eliminates the need for wired sensors, simplifying access to harsh, remote or hard-to-reach environments. The ValProbe system is ideal for applications where high measurement accuracy and regulatory compliance are priorities:

    • Pharmaceutical processing
    • Sterilization of medical devices
    • Food processing
    • Environmental monitoring

    Integrated data processing and reporting capabilities extend the operating convenience of ValProbe systems beyond mere data acquisition. The ValProbe system performs calculations and generates user-defined custom reports for up to 200 sensors at a time. Graph reports can include many sensors and limits for easy review of study data. User calibration and verification is a quick and easy process with the new CTR-40 temperature reference.

    ValProbe meets the requirements of FDA 21 CFR Part 11 for electronic signatures and registers and complies with ISO-17665, EN285 and HTM-2010 for saturated steam sterilization.

    For extremely low temperatures, see the Kaye ValProbe® Cryo Logger with a temperature range of -85 ° C to 140 ° C

    {tabCaracterísticas}

    • La tecnología RTD ofrece una precisión de medición sin igual en un amplio rango operativo
    • El sistema descarga hasta 10 sondas simultáneamente.
    • Fácilmente definido recopilación de datos basada en el ciclo, cálculo e informes de hasta 200 sensores
    • Permite el cumplimiento de la normativa FDA 21 CFR Parte 11
    • Funciona desde -45 ° C hasta 140 ° C, y hasta 10 BAR absolutos
    • Diseñado para una fácil calibración en el sitio
    • Indicador de duración de la batería y batería reemplazable en el campo
    • Disponible como un kit de inicio de validación económica completo con software, lector único, dos registradores de temperatura y uno de presión (P / N V2543).
    • Capacidad de informes mejorada con hasta 25 grupos programables por el usuario.

    {tab Aplicaciones}

    El sistema ValProbe es ideal para aplicaciones donde la alta precisión de medición y el cumplimiento normativo son prioridades:

    • Esterilizadores
    • Hornos
    • Liofilizadores
    • Cámaras de estabilidad
    • Cámaras de temperatura
    • Frigorificos, Congeladores
    • Almacenes

    {tab Especificaciones}

    Kaye ValProbe Generic Specifications

    Reader 2 Dimensions

    320mm x 155mm x 60mm

    (12,60” x 6,10” x 2,36”)

    Logger Dimensions

    Height: 48mm / Diameter: 36mm

    (Height: 1,9’’ / Diameter: 1,4’’)

    Logger Material

    Stainless Steel 316L and Peek

    Battery           

    Field replaceable – 3.6V Lithium

    Sampling Rate

    1 sec to 12 hours

    Data Storage             

    10,000 Samples retained in non-volatile memory

    Real-Time Clock Accuracy  

    < 15sec / day

    Calibration   

    NVLAP/DAkkS Calibration

    Verification

    Automated User

    Verification capability

    Sensing Element

    Precision Platinum RTD

    Environmental Temperature

    -85°C/-45°C to 140°C

    Environmental Pressure

    0 – 10 bar absolute

    Environmental Humidity

    0 – 100% condensing

    Regulatory Compliance

    UL and CE

    Kaye ValProbe Temperature Loggers Specifications           

    Specifications

    Rigid

    Flexible

    Bendable

    Surface

    Sensor Type  

    Single Sensor

    Single Sensor

    Single and Dual SENSOR

    Sensor

    Sensor Length   

    1.5, 3, 6, 9” inches

    (38, 76, 152, 229mm)

    40” inches

    (1000mm)

    12, 24, 36” inches

    (305, 610, 915mm)

    -

    Tip Diameter

    0.118” (3mm)   

    0.095” (2.4mm)

    0.095” (2.4mm)

    1.26” (32mm)

    Measurement Range

    -85°C / -45°C to 140°C

    -85°C / -45°C to 140°C

    -85°C / -45°C to 400°C

    -85°C to 140°C

    Accuracy  

    -85°C to 0°C, ±0.25°C / Cryo VP;

    -45°C to 0°C, ±0.25°C / Standard VP;

    0°C to 140°C, ±0.1°C

    -85°C to 0°C, ±0.25°C / Cryo VP;

    -45°C to 0°C, ±0.25°C / Standard VP;

    0°C to 140°C, ±0.1°C

    ±0.25°C from -85°C to 0°C / Cryo VP; 

    ±0.25°C from -45°C to 0°C / Standard VP;

    ±0.1°C from 0°C to 140°C;

    ±0.2°C from 140°C to 250°C;

    ±0.5°C from 250°C to 360°C

    -85°C to 0°C, ±0.25°C;

    0°C to 140°C, ±0.1°C

    Kaye ValProbe Pressure, Temperature and Humidity Logger Specifications            

    Specifications    

    Pressure & Temp

    Humidity & Temp

    Environmental Temperature

    0°C to 140°C

    0°C to 90°C 

    Sensor Type

    Single P/T Sensor

    RH: EMD4000

    Tip Diameter  

    1/4” NPT
    Connection Fitting

    Diameter 18mm (0,71”)

    Length 35mm (1,38”)

    Measurement Range

    0°C to 140°C /

    0 – 5bar abs.

    25% to 85% RH

    Accuracy

    0°C to 120°C; ±25mb

    120°C to 135°C; ±10mb

    135°C to 140°C; ±25mb

    0°C to 140°C; ±0.1°C

    25% to 85% RH; ± 2% RH (at 25°C to 40°C)

    0°C to 90°C; ±0.1°C

    {tab Descargas}

    English

    Portuguese

    Chinese

    {tab Video}

    {/ tabs}

    ico pointerKAYE HEADQUARTERS

    Amphenol Advanced Sensors Germany GmbH 
    Sinsheimer Strasse 6
    D-75179 Pforzheim

    T +49 (0) 7231-14 335 0
    F +49 (0) 7231-14335 29
    kaye-ccpf@amphenol-sensors.com  
    www.kayeinstruments.com 

    KAYE Americas 

    Amphenol Thermometrics, Inc.
    967 Windfall Road
    St. Marys, PA 15857
    Tel: +1(814) 834-9140
    Fax: +(814) 781-7969
    kaye-us@amphenol-sensors.com 


    Company Details  |  Disclaimer/ Terms of Use  |  Privacy Policy

    ico pointer  QUICK CONTACT

    9 + 1 =
    7 + 7 =
    9 + 1 =
    9 + 8 =
    9 + 4 =

    ico alert ALERTA DE NOTICIAS DE KAYE  

    Manténgase informado sobre nuestros últimos productos y actualizaciones importantes de software.

    SUSCRIBETE PARA RECIBIR NOTICIAS