The battery life of the Kaye ValProbe RT logger depends on several factors, including the sampling rate, ambient temperature, and whether RF transmission is enabled. Under normal conditions and with a standard configuration, the logger—equipped with a field-replaceable 3.6 V Lithium Thionyl Chloride ½ AA battery—can provide up to 5,000 hours of operation.
The ValProbe RT software continuously monitors battery status and estimates remaining life with an accuracy of ±10%, helping prevent unexpected interruptions during studies. It calculates battery life based on energy output over the temperature range. However, due to changes in battery chemistry, remaining battery capacity estimated at room temperature may be reduced when used in colder environments.
For the highest accuracy, battery life estimates are most reliable when the logger is used exclusively in either room temperature or low-temperature applications—not in a mix of both.