The total system error — also known as total measurement uncertainty — is a combined value that accounts for all potential sources of deviation within a measurement or validation setup. In regulated environments such as pharmaceutical, biotechnology, or food production, accurate calculation of total system error is essential for ensuring data reliability and compliance with standards (e.g. FDA, GMP, or EN ISO 17025).
To calculate the total system error (TSE) accurately, you typically combine multiple error sources using the root-sum-square (RSS) method:
Total System Error (TSE) = √(E₁² + E₂² + E₃² + ... + En²)
Where:
• E₁, E₂, E₃, etc. represent individual errors from different system components, such as:
o Sensor accuracy (e.g. thermocouple, RTD, wireless logger)
o Logger resolution and accuracy
o Reference standard uncertainty (e.g. IRTD calibration)
o Calibration equipment uncertainty (e.g. dry block or liquid bath)
o Environmental or setup conditions (e.g. thermal gradients, immersion depth)
Example (for Kaye systems such as AVS Validator or ValProbe RT):
Imagine the following typical tolerances:
• Sensor accuracy: ±0.2 °C
• Logger/system accuracy: ±0.1 °C
• Calibration reference (e.g. IRTD): ±0.025 °C
Then: TSE = √(0.2² + 0.1² + 0.025²) = √(0.04 + 0.01 + 0.000625) = √0.050625 ≈ ±0.225 °C
This total system uncertainty represents the combined effect of all included sources, assuming they are independent and normally distributed.
In our Kaye product line we provide documented system accuracy and uncertainty levels in the technical specifications and calibration certificates. Our systems are designed to meet demanding regulatory validation needs, and the total system error is often already defined and validated during IQ/OQ processes.