The Pass/Fail tool in Kaye validation software — used with both the AVS Validator and ValProbe RT systems — allows users to automatically evaluate measurement data against predefined acceptance criteria. This feature streamlines the validation process by identifying whether individual channels, sensors, or entire qualification cycles meet the required specifications for temperature, pressure, humidity, or lethality (F₀) parameters.
Key functions include:
-Defining upper and lower tolerance limits for each measured parameter
-Automatically flagging out-of-spec data points or cycles
-Generating clear, audit-ready reports showing compliance or deviations
-Supporting regulatory and quality requirements (e.g., GMP, FDA, EU Annex 1)
This tool is especially beneficial when performing thermal mapping in autoclaves, stability chambers, warehouses, or cleanrooms — enabling fast, consistent, and compliant review of measured data without the need for manual calculations or searching.