English

Can F₀ be calculated after the study is completed if the lethality calculation parameter was not configured during setup?

Yes, the Kaye Common Reporting Tool (CRT) allows you to define a new cycle and apply custom settings after the study has been completed. This means you can retroactively calculate the F₀ (lethality value) even if the lethality calculation parameter was not initially configured during the setup phase.
CRT provides flexible reporting tools to ensure you can generate all critical validation data — including F₀ — in compliance with regulatory requirements.

If you need assistance creating a new cycle or setting F₀ parameters in CRT, please contact Kaye technical support or one of our certified service partners.