What are the temperature limitations for battery-operated logger systems, and why do these limits exist?

The temperature limitations of Kaye ValProbe RT battery-powered loggers are primarily defined by the characteristics of the lithium batteries used.
For high-temperature applications above +140 °C, a thermal protective enclosure (insulated canister) must be used. This allows operation in environments up to +400 °C for a limited period of time, depending on the insulation type and setup.
For low-temperature conditions below –40 °C, a battery extension can be used to optimize performance. Additionally, when the logger operates below –40 °C, the real-time RF transmission is automatically disabled to preserve battery life. During this phase, all measurement data is stored in the internal memory. Once the logger returns to above –40 °C, it will automatically reconnect and transmit all stored data to the base station with a complete time stamp.