The full calibration and adjustment process of an IRTD-400 reference thermometer takes approximately 4 working days. This duration is necessary to ensure extremely low measurement uncertainty of ±0.025 °C across the entire temperature range of -196 °C to +420 °C.
The following key steps are included:
As-Found Report
An initial measurement is performed across the full range (-196 °C to +420 °C) to document the sensor’s condition before any adjustment.
Annealing Cycle (60 hours at 420 °C)
This is a critical thermal treatment step that stabilizes the sensor’s performance in the long term. It is often skipped by some calibration labs due to the long duration — however, it is essential to meet the required accuracy and reliability standards.
Why annealing?
Reduces internal mechanical stress in sensor materials
Improves long-term measurement stability
Minimizes sensor drift under thermal cycling
Enhances reproducibility of measurements
Restores and optimizes the crystal lattice structure, especially in platinum RTDs
Ensures consistent electrical and thermal behavior
Adjustment Phase
The IRTD is adjusted at several fixed temperature points from -196 °C to +420 °C to align it to known standards.
Calibration Cycle
After adjustment, the IRTD undergoes a final full-range calibration over 7 temperature points. This step verifies the adjustment and ensures conformity to its specification. The final result is documented in a traceable calibration certificate.
This meticulous process is part of what makes the Kaye IRTD-400 one of the most reliable and stable reference sensors available — fully compliant with the rigorous standards required in GMP/FDA-regulated environments.