English

Can I perform in-house calibration of my ValProbe® RT loggers for temperature, pressure, and relative humidity?

Yes, as part of a sound risk management process, it is recommended to verify the accuracy of data loggers periodically between annual recalibrations. The ValProbe® and ValProbe® RT software include a built-in feature that supports this type of intermediate verification.

However, please note that this verification process does not include adjustment of deviations (no sensor re-alignment is performed). It serves only to confirm continued accuracy. A full calibration with adjustment can only be performed by the manufacturer.

To ensure valid results during in-house verification, appropriate reference equipment such as certified temperature baths, dry block calibrators, or traceable RH/pressure standards must be used.