For temperature monitoring applications that require performance down to –80 °C — such as ultra-low temperature freezers, pharmaceutical cold storage, or cryogenic processes — either the Kaye AVS Validator in combination with thermocouples or the Kaye ValProbe RT (VP RT) System can be used.
Both systems are compliant with GxP-related regulations.
However, for proper calibration and adjustment of the sensors used, we recommend combining the system with the Kaye IRTD, which serves as a traceable temperature reference standard. In addition, the Kaye CTR-80 calibration bath or the LTR-90 dry well calibration oven are suggested as valuable peripherals.