English

Can the AVS be used for validating parameters other than temperature and pressure?

Yes, the AVS is designed to support a wide range of validation applications beyond just temperature and pressure. It is compatible with all linear analog signals with standard outputs of 0–10 V or 4–20 mA. This allows the integration of various sensor types, including thermocouples, Pt-100 RTDs, relative humidity (RH) sensors, CO₂ sensors, and other transducers. The flexible input module architecture of the AVS enables accurate monitoring and documentation of numerous environmental parameters, making it suitable for complex validation requirements in GxP-regulated environments